Dual-fan system

· Unique design of centrifugal fan+axial flow fan with dual-winding speed sensor realizes real-time detection of air downflow rate and inflow rate, smart adjustment of rotational speed, to maintain airflow at stable velocity, bring better protection to experimenters and samples and make MBH be wildly used in various experimental environments.

 High-efficiency HEPA

· High-efficiency HEPA with low loss of pressure; filtration efficiency of particles(>0.3um) is up to 99.995%.

 User-friendly design

· Integrated structure: Internal side wall and rear wall are designed as integrated structure to prevent leakage.
· Multi alarms for abnormal airflow, abnormal door location, sensor failure, etc.
· Fixture feet without exposed thread, height adjustment structure with casters to facilitate cabinet removing.
· With self-cleaning function after power-on.
 

 CLASS Ⅱ TYPE A2

External Dimensions 1060×806×2216 mm
Internal Dimensions 960×630×660 mm
Internal Work Area,Space 0.6㎡
Downflow Velocity 0.32m/s
Inflow Velocity 0.53m/s
Volume 675m³/h
Inflow Volume 458m³/h
Downflow Filter HEPA H14 99.995%@MPPS
Inflow Filter HEPA H14 99.995%@MPPS
ULPA Filter(Optional) ULPA U15 99.999%@MPPS
Valve(Optional) Air/Water Valve

 Biosafety Protection Test

Personnel Protection Test1~8×108 (three times in succession) KI-Discus containment and microbiological testing is performed
Product Protection Test1~8×106(three times in succession) ≤5CFU
Cross-contamination Test1~8×104(three times in succession) ≤2CFU
Average Light intensity(Lux) ≥900(220V)
UV radiation intensity >400mW/sqm
Noise <65dB
Excellent light distribution YES
RMS <5μm
Main Body 1.5mm Galvanized Steel
Work Zone 1.5mm Stainless Steel, type 304
Front Window 10°,6mm Tempered Glass, UV Safety, Anti UV
Net Weight 260kg
Shipping Weight 295kg
Shipping Dimensions Maximum(W×D×H) 1140×1103×1840
Shipping Volume, Maximum(m³) 2.31

 220V~50Hz

Cabinet Full Load Amp(FLA) 1.3A
Fuses 10A
Cabinet Nominal Power 204W
Optional Outlets FLA 5A
Total Cabinet FLA 6.3A

 Alarm

Airflow velocity sensor fault, low/high airflow velocity fault, airflow blower fault, pressure alarm, filter/UV lifespan alarm, door position alarm
Filter Standard
 
HEPA H14 accordance with EN 1822-4: efficiency of 99.995%@MPPS 
ULPA U15 accordance with EN 1822-4: efficiency of 99.999%@MPPS
Certification Standard Certification Standard

REQUEST A QUOTE

SERVICE HOTLINE

+86 18640846040

COPYRIGHT @ 2023 冰洋生物科技(大连)有限司
辽ICP备2004985号
辽网安备 21021702000472号