· Strictly follow the design standard YY1007, in compliance with GB150 for requirements on pressure vessel.
· Stainless steel(in compliance with GB/T24511)plate with 3mm in width improves safety performance vessel.
· Dual cooling radiating system realizes rapid pull-down.
· 2 standard cooling fans at the left rear and left front of chamber reduce the cooling time.
· HEPA filter is optional for particles larger than 0.2um, filtering effect> 99.5%.
· Lid opening at low temperature and exhaust under high pressure: double dry burning-resistant protection, safe interlock and structure.
· Users can input the actual altitude and the system will automatically calculate the matching boiling point and running parameters.
· Press the operation key, and the sterilizer will start up with previous running parameters, which is convenient efficient and more applicable for bulk repetitive operations.
· Malfunction display, easy tmmo find the reasons.
Specifications |
|
External Dimensions (WxD x H) | 600×620×820mm |
Internal Dimensions (Wx D xH) | Φ370×610mm |
Volume | 55L |
Net weight | 93kg |
Performance |
|
Sterilization temperature | 105~135℃ |
Dissolution temperature | 60~110℃ |
Dissolution temperature | 45~60℃ |
Altitude settings | 0 to 4000m |
Noise Level | <65db |
Construction |
|
Exterior Material | Painted Steel |
Interior Material | 304 |
Lid | Resin anti-scald lid |
Lid lock | Mechanical lock |
Casters | 4 casters |
Controller |
|
Controller | Microprocessor control |
Display | 5 inch touch screen |
Adjustment accuracy | 0.1℃ |
Design pressure | 0.3 Mpa |
Safety valve discharge pressure | 0.28 Mpa |
Pressure gauge range | 0~0.4Mpa |
Temperature Sensor | PT100 |
Sterilization procedure |
1. Quick sterilization 2. Sterilization and insulation 3. Liquid sterilization 4. Dissolution insulation 5. Preheat sterilization |
Alarm(V=Visual Alarm, B=Buzzer Alarm, R=Remote Alarm) |
|
Temperature alarm | V-B |
Pressure alarm | V-B |
Water level alarm | V-B |
Door lock alarm | V-B |
Heating abnormality alarm | V-B |
Main control panel alarm | V-B |
Electrical |
|
Power Supply | 220~240V |
Frequency | 50/60Hz |
Rated input power | 2850W~3350W |
Communication systems |
|
Standard interface | —— |
Optional interface | USB、485 Modbus |
Standard Accessories |
|
Standard |
1. Drainage pipe x1; 2. Tray x1; 3. Sterilization basket x2(MVS-64-E)/sterilization basket x3(MVS-84-E); 4. Cooling bucket; 5. Condensate box; 6. Manual |
Optional Accessories |
|
Optional |
Dot matrix printer or thermal paper printer USB、485 Modbus Tempertature sensor(For liquid sterilization) |
Quality Management System |
|
Certification | 2014/68/EU, ASME SA-240 304 |
CATALOGS
REQUEST A QUOTE
SERVICE HOTLINE
+86 18640846040