Parameter

External Dimensions (WxD x H) W626mm×D874mm×H899mm
Internal Dimensions (Wx D xH) 490 × 525× 666 mm
Volume 170L
Net weight 108kg

 Performance

Temperature control  range AT+5~50℃
Temperature Uniformity ±0.25℃(AT:22°C、Settings:37°C、CO2:5%、no load )
Temperature Fluctuation ±0.1°C(AT:22°C、Settings:37°C、CO2:5%、no load )
CO2 concentration control method PID
Display 7 inch LCD
CO2 sensor IR sensor
CO2 Control Range 0 ~ 20 %
CO2 Fluctuation ±0.15 %(AT:22°C、Settings:37°C、CO2:5%、no load )
Ambient temperature 5~35℃
Heating method Direct Heat & Air Jacket System
Humidification method Natural evaporation of water in the tray 
Humidification method 95±5%
Noise Level  (db) ≤45dB(A)
Sterilization method Hot Air Sterilization(180°C)+UV Lamp 4Wx1
Circulation method Air stirring type
O2 concentration control method None
O2 sensor None
O2 Control Range None
O2 Fluctuation None
Pull up and pull down   
Temperature and carbon dioxide concentration recovery curve 30s +60s after opening the door
 

 Construction

Exterior Material Painted Steel
Interior Material SUS304 steel
Insulation materials Rock wool
Heating wire Silicone heating wire 1186W
Outer Doors/Lid Painted Steel 
Outer Door Lock Optional
Inner Doors/Lid Tempered glass
Tray material SUS304
Tray deminsion 469*444*9mm
Tray number 4pcs
Filter 0.3um efficiency: more than 99.97%
CO2 connection port Pagoda connector
CO2 connection port ——

 Controller

Temp sensor Dual IR sensor
HS controller PWM
Accuracy value 0.1
UV lamp 4W×1
Access Port 1
Access Port Position Back 
Access Port Diameter 30mm
Casters Adjustable feet

 Alarm(V=Visual Alarm, B=Buzzer Alarm, R=Remote Alarm)

CO2 concentration exceeds upper limit alarm V-B-R
CO2 concentration exceeds lower limit alarm V-B-R
Door open alarm V-B-R
Alarm for empty CO2 cylinder or CO2 solenoid valve failure V-B-R
UV lamp alarm V-B-R
Ambient temperature sensor failure V-B-R
High temperature protection sensor failure in the box V-B-R
High temperature alarm inside the box V-B-R
Low temperature alarm inside the box V-B-R
High temperature protection alarm inside the box V-B-R

 Electrical

Power Supply(V) 220~240V
Frequency(Hz) 50
Rated input power 1200W
Power consumption ——

 Standard Accessories

Accessories
 
Tray×4、gas pipe×1、gas pipe clip×2、humidification tray×1、stacking board×1、wear pad×2、M4 screw×4、M5 gasket×4、M8 gasket×4

 Options

Options Platform trolley
  Electronic lock

 Communication Systems

Standard port RS485
Optional port USB

 Quality Management System

Certification ——
      
 

REQUEST A QUOTE

SERVICE HOTLINE

+86 18640846040

COPYRIGHT @ 2023 冰洋生物科技(大连)有限司
辽ICP备2004985号
辽网安备 21021702000472号
联网药信息服务资格证书编号: 非经营件2020-0035